Skip to content

Error Codes

The GE command returns the current error code. If no error occurred, it is 0. When an error leads to switching off the laser, GE is set accordingly. The next time the laser is turned on successfully, GE is reset to 0.

A list of error codes and their causes follows:

error codecause
0no error, everything ok
1interlock open
2laser compliance voltage not OK or no laser connected
3internal supply voltage not OK
4laser temperature sensor open
5crystal temperature sensor open
6laser temperature exceeds upper limit (1SLU)
7laser temperature lower than lower limit ( 1SLL)
8laser short-circuit or no laser connected
9device temperature (GT) too high
10laser temperature exceeds maximum laser temperature (LTM)
11crystal temperature exceeds upper limit (2SLU)
12crystal temperature lower than lower limit (2SLL)
16laser current greater than maximum current limit (LCLM
17current error
18total power limit exceeded